Perform ance C heck and Functional Verification Procedure
2440 Service
WARNING
|
DO NOT connect the 10X Probe ground lead to
the AC (line) power source when performing
this step.
b.
Connect a 10X Probe to the CH 1 OR X input connector
and connect the 10X Probe tip to an AC (line) source.
c.
Using the TRIGGER LEVEL control, VERIFY that the
display can be triggered on the positive going slope of
the AC waveform for the + (plus) selection of the
SLOPE button and on the negative going slope for the -
(minus) selection of the SLOPE button.
NOTE
The Trigger Point Indicator, a small “T" riding on
the displayed waveform, indicates the point on
which the instrument is triggered for the
displayed waveform.
d.
Disconnect the test setup.
7. Verify A and B Trigger Position Function.
a.
Recall the Initial Front-Panel Setup, labeled “FPNL”
(see step i in “INITIAL SETUP" at the start of this
procedure). Make the following changes to the
front-panel setup:
Select: VERTICAL M O DE
Set:
CH 2
Off
Set:
CH 1 VOLTS/DIV
IV
b.
Connect the MARKER O U T signal of the Tim e Mark
Generator to the CH 1 OR X input connector through a
50
n
Coaxial Cable.
c.
Set the Time Mark Generator’s marker period to 1 ms.
d.
Position the start of the display to the extreme left
graticule line using the HORIZONTAL POSITION
control.
e.
Select TRIG POSITION and set Vs on.
f.
V E R IF Y -T h a t the Trigger Point Indicator (a “T ”
symbol) is positioned on a time marker approximately
2.5 divisions to the right of the
extreme
left
graticule line.
g.
Set the TRIGG ER POSITION menu to V4 and verify that
the Trigger Point Indicator moves to a time marker that
is approximately at center screen.
h.
Use the HORIZONTAL POSITION control to position
the time marker with superimposed Trigger Point
Indicator to the extreme left graticule line.
i.
Set the TRIGG ER POSITION menu to V
2
and verify that
the Trigger Point Indicator moves to a time marker that
is approximately at center screen
j.
Use the HORIZONTAL POSITION control to position
the time marker with superimposed Trigger Point
Indicator to the extreme left graticule line.
k.
Set the TRIGG ER POSITION menu to 3/ 4 and verify that
the Trigger Point Indicator moves to a time marker that
is approximately at center screen.
l
.
Set the TRIGG ER POSITION menu to 7/ 8 and verify that
the Trigger Point Indicator is positioned on a time
marker approximately 2 .5 divisions to the right of the
center graticule line.
m. Press A /B TRIG to select the B Trigger System and set
the HORIZONTAL mode to B. Use the TRIGGER LEVEL
control to trigger the display as required.
n.
Repeat parts d through k to check the B TRIGGER
P O SITIO N function.
o.
Disconnect the test setup.
4-23
Содержание 2440
Страница 4: ......
Страница 12: ......
Страница 37: ...Specification 2440 Service Dimensions are in inches mm 4918 39 1 25 Figure 1 1 Dimensional drawing ...
Страница 38: ......
Страница 91: ...Theory of Operation 2440 Service 3 49 Figure 3 5 Simplified CCD architecture ...
Страница 120: ...Theory of Operation 2440 Service Figure 3 10 DC Restorer 3 78 ...
Страница 123: ...Theory of Operation 2440 Service SOURCE ACCEPTOR 4 9 1 7 1 2 3 81 Figure 3 11 GPIB data flow diagram ...
Страница 130: ...Theory of Operation 2440 Service 3 88 ...
Страница 136: ...Theory of Operation 2440 Service Figure 3 15 PWM Regulator and Inverter 3 94 ...
Страница 138: ...Theory of Operation 2440 Service Figure 3 16 PWM switching waveforms 3 96 ...
Страница 208: ......
Страница 322: ......
Страница 419: ......
Страница 423: ...2440 Service Figure 9 4b 2440 Block diagram part 2 6330 27 BLOCK DIAGRAM PART 2 ...
Страница 427: ......
Страница 428: ...2440 Service WAVEFORMS FOR DIAGRAM 1 6 2 8 5 3 2 ...
Страница 429: ......
Страница 432: ...2440 Service WAVEFORMS FOR DIAGRAM 2 BUS ISOLATED 6 6 0 3 4 9 ...
Страница 433: ...WAVEFORMS FOR DIAGRAM 2 ...
Страница 439: ......
Страница 445: ......
Страница 446: ...2440 Service WAVEFORMS FOR DIAGRAM 7 I I f f f NtITOT CHI r K I It MVIHi TRIGGER ON FALLING EDGE OF U832 PIN 8 MORE ...
Страница 450: ...2440 Service WAVEFORMS FOR DIAGRAM 8 6603 27 ...
Страница 455: ...W A V EFO R M S FOR D IAG RAM 11 WAVEFORMS FOR DIAGRAM 11 6603 29 ...
Страница 458: ...WAVEFORMS FOR DIAGRAM 12 ...
Страница 463: ...WAVEFORMS FOR DIAGRAM 14 ...
Страница 468: ...2440 Service W A VE FO R M S FOR DIAG RAM 17 TEST SCOPE TRIGGERED ON 126 6330 46 WAVEFORMS FOR DIAGRAM 17 ...
Страница 471: ... C N i u i t e x r 4 j u J l i 1 i 6 28 5 44 WAVEFORMS FOR DIAGRAM 18 ...
Страница 476: ...WAVEFORMS FOR DIAGRAM 20 ...
Страница 490: ...YES ...
Страница 492: ......
Страница 494: ......
Страница 499: ...YES ...
Страница 501: ......
Страница 502: ...2440 Service FIG 1 C A B IN E T ...
Страница 503: ......
Страница 510: ...2440 Service SISSVHO Z Old ...
Страница 511: ......
Страница 512: ...A12 2440 Service F IG 3 C IR C U IT BOARDS ...
Страница 513: ......
Страница 517: ...10 16 R E V J U L 1993 Replaceable Mechanical Parts ...
Страница 518: ...2440 Service ...
Страница 519: ...FIG 4 L V H V PWR SUPPLY ...
Страница 520: ...FIG 5 ACCESSORIES 2440 Service ...
Страница 521: ......
Страница 523: ......
Страница 529: ......
Страница 530: ......
Страница 531: ......