9. Check Square-Wave Flatness (Video Option
05 only).
a.
Recall the Initial Front-Panel Setup, labeled “FPNL”
(see step I in “INITIAL S E T U P ” at the start of this
procedure). Make the following changes to the front-
panel setup:
Select: CH 1 C O U P L IN G /IN V E R T
Set:
5 0 0 O N |O FF
OFF
Select: CH 2 C O U P L IN G /IN V E R T
Set:
5 0 0 ON | OFF
OFF
Set:
CH 1 V O LTS/D IV
20 0 mV
CH 2 V O LTS/D IV
50 mV
A S E C /D IV
2 ms
Select: VER TIC A L M O D E
Set:
CH 2
O ff
b.
Connect the fast-rise, positive going square-wave
output of the Calibration Generator to the CH 1 OR X
input connector via a 50 Q Coaxial Cable and a 50 O
Termination. The square wave should step from -1 V to
o v .
c.
Set the generator to produce a 60-H z, five-division
display and use the CH 1 POSITION control to center
the display as required.
d.
Set the CH 1 VOLTS/DIV control to 50 mV.
e.
C H E C K -D is p la y front-corner aberrations are within
1% (0.2 division or less). Exclude the first 20 ns
immediately following the positive going transition
from the measurement.
f.
Select VERTICAL M O DE and set CH 2 and CH 1 off.
g.
Move the 50 O Coaxial Cable from the CH 1 0 R X input
connector to the CH 2 OR Y input connector.
h.
C H E C K —Display front-corner aberrations are within
1% (0.2 division or less). Exclude the first 20 ns
immediately following the positive going transition
from the measurement.
I.
Set the CH 2 VOLTS/DIV control to 5 mV.
j.
Install a 10X attenuator between the 50 0 Coaxial
Cable and the 50 O Termination, and reconnect the
setup.
k.
C H E C K -D is p la y front-corner aberrations are within
1% (0.2 division or less). Exclude the first 20 ns
immediately following the positive going transition
from the measurement.
l
.
Select VERTICAL M O DE and set the CH 2 VOLTS/DIV
control to 50 mV, then set CH 1 on and CH 2 off.
Perform ance C heck and Functional Verification Procedure
2440 Service
m. Move the test setup from the CH 2 OR Y input connector
to the CH 1 OR X input connector. Set the CH 1
VOLTS/DIV control to 5 mV.
n.
CHECK —Display front-corner aberrations are within
1% (0.2 division or less). Exclude the first 20 ns
immediately following the positive going transition
from the measurement.
o.
Set the CH 1 VOLTS/DIV control to 200 mV and set the
A S E C /D IV control to 10 |xs.
p.
Remove the 10X attenuator and reconnect the
test setup.
q.
Set the Calibration Generator to produce a 15-kHz,
5-division display.
r.
Repeat parts d through n to check square-wave flatness
at 15 kHz.
s.
Disconnect test setup.
10. Check Frequency Response Flatness (FULL
and 20 M Hz BANDWIDTH Modes) (Video Option
05 only).
a.
Recall the Initial Front-Panel Setup, labeled “FPNL”
(see step i in “INITIAL SETUP" at the start of this
procedure). M ake the following changes to the front-
panel setup:
Select:
CH 1 COUPLIN G /IN VER T
Set:
5 0 0 O N | OFF
OFF
Select:
C H 2 COUPLIN G /IN VER T
Set:
5 0 0 O N | OFF
OFF
4-35
Содержание 2440
Страница 4: ......
Страница 12: ......
Страница 37: ...Specification 2440 Service Dimensions are in inches mm 4918 39 1 25 Figure 1 1 Dimensional drawing ...
Страница 38: ......
Страница 91: ...Theory of Operation 2440 Service 3 49 Figure 3 5 Simplified CCD architecture ...
Страница 120: ...Theory of Operation 2440 Service Figure 3 10 DC Restorer 3 78 ...
Страница 123: ...Theory of Operation 2440 Service SOURCE ACCEPTOR 4 9 1 7 1 2 3 81 Figure 3 11 GPIB data flow diagram ...
Страница 130: ...Theory of Operation 2440 Service 3 88 ...
Страница 136: ...Theory of Operation 2440 Service Figure 3 15 PWM Regulator and Inverter 3 94 ...
Страница 138: ...Theory of Operation 2440 Service Figure 3 16 PWM switching waveforms 3 96 ...
Страница 208: ......
Страница 322: ......
Страница 419: ......
Страница 423: ...2440 Service Figure 9 4b 2440 Block diagram part 2 6330 27 BLOCK DIAGRAM PART 2 ...
Страница 427: ......
Страница 428: ...2440 Service WAVEFORMS FOR DIAGRAM 1 6 2 8 5 3 2 ...
Страница 429: ......
Страница 432: ...2440 Service WAVEFORMS FOR DIAGRAM 2 BUS ISOLATED 6 6 0 3 4 9 ...
Страница 433: ...WAVEFORMS FOR DIAGRAM 2 ...
Страница 439: ......
Страница 445: ......
Страница 446: ...2440 Service WAVEFORMS FOR DIAGRAM 7 I I f f f NtITOT CHI r K I It MVIHi TRIGGER ON FALLING EDGE OF U832 PIN 8 MORE ...
Страница 450: ...2440 Service WAVEFORMS FOR DIAGRAM 8 6603 27 ...
Страница 455: ...W A V EFO R M S FOR D IAG RAM 11 WAVEFORMS FOR DIAGRAM 11 6603 29 ...
Страница 458: ...WAVEFORMS FOR DIAGRAM 12 ...
Страница 463: ...WAVEFORMS FOR DIAGRAM 14 ...
Страница 468: ...2440 Service W A VE FO R M S FOR DIAG RAM 17 TEST SCOPE TRIGGERED ON 126 6330 46 WAVEFORMS FOR DIAGRAM 17 ...
Страница 471: ... C N i u i t e x r 4 j u J l i 1 i 6 28 5 44 WAVEFORMS FOR DIAGRAM 18 ...
Страница 476: ...WAVEFORMS FOR DIAGRAM 20 ...
Страница 490: ...YES ...
Страница 492: ......
Страница 494: ......
Страница 499: ...YES ...
Страница 501: ......
Страница 502: ...2440 Service FIG 1 C A B IN E T ...
Страница 503: ......
Страница 510: ...2440 Service SISSVHO Z Old ...
Страница 511: ......
Страница 512: ...A12 2440 Service F IG 3 C IR C U IT BOARDS ...
Страница 513: ......
Страница 517: ...10 16 R E V J U L 1993 Replaceable Mechanical Parts ...
Страница 518: ...2440 Service ...
Страница 519: ...FIG 4 L V H V PWR SUPPLY ...
Страница 520: ...FIG 5 ACCESSORIES 2440 Service ...
Страница 521: ......
Страница 523: ......
Страница 529: ......
Страница 530: ......
Страница 531: ......