50
SV106A User Manual
5.11.1 Example of timer function execution
The
Timer
function is used to programme the instrument to switch on at the
desired time and perform the measurements with the parameters set in the
Measurement
sub-list.
Let us assume that the user wants to switch on the instrument on the 1
st
of
January, at 20:30, measure the vibration during 10 seconds without using
logger and save the results in a file named @R1.
In order to do this the user has to set the parameters of the
Timer
function
(
path: Menu / Measurement / Timer
), the measurement parameters (
path:
Menu / Measurement / General Settings
), activate the
Auto Save
function
(
path: Menu / File / Save Options
), name the file (the
File Name
window is
opened after switching on the
Auto Save
function) and finally
– switch off
the instrument.
The instrument will be switched on the 1
st
of January at 20:30 and will be
warmed up for the period of 60 seconds decrementing the counter visible
on the display by one after each second.
After warming up the instrument and the pre-set
Start Delay
time, the
measurements will be performed for a period of ten seconds. Then, the
results will be saved in the previously named file and finally
– the instrument
will switch off.