SV106A User Manual
Activation the event recording
The
Event Recording
position enables the
user to activate the event recording
function. This position doesn’t appear in the
Simple Mode
(
path:
<Menu>
/
Auxiliary Setup / Instrument Mode
).
If
Event Recording
is switched off, then the
Event Recording
position
in
the
Data Logging
list will be not active.
=>
Note: The Event Recording function is optional and should be unlocked by entering the activation
code in the text editor screen, which is opened after first attempt to select it. Once unlocked, this
option will be ready to use permanently.
In case the SV 105AF with contact force sensor is used, the
Force Level+
parameter appears in the list. It allows to set the threshold above which the
instrument recognises contact force and assess the time of contact with the
tool.
Level+
time will be displayed during measurement in Dosimeter
presentation mode.
5.2 Setting the parameters for dose measurements
– HAV/WBV Dosimeter
The
HAV/WBV Dosimeter
list enables the user to set up the parameters
for vibration dose measurements, like: exposure period, type of
measurement (whole body or hand arm) performed in channels 1-3 and 4-
6, vibration action limits used for some standards (
U.K.
,
Italy
,
Poland
,
France
,
Germany China
,
Brazil
), as well as specific limits defined by the
user (
User
).
Setting the measurement type for channels 1-3 and 4-6
Positions
1-3 Dosimeter
and
4-6 Dosimeter
enable the user to set the
desired type of the measurement, performed with the use of channels 1,2,3
and 4,5,6
– hand-arm (
HAV
) or whole-body (
WBV
) vibration.
Setting the exposure time
The
Exposure Time
enables the user to set the desired value of the
exposure time that is used for the calculation of the
HAV/WBV Dose
results. The
Exposure Time
values are within the range [
00h01
,
24h00
].