![Svantek SV 106A Скачать руководство пользователя страница 42](http://html1.mh-extra.com/html/svantek/sv-106a/sv-106a_user-manual_1409261042.webp)
42
SV106A User Manual
The
Trigger
position enables the user to
select the trigger type:
Level -
,
Level +
,
Slope -
,
Slope +
,
Gradient -
or
Gr
.
In each interval of the measurement, defined by
Trig. Step
, the triggering condition is checked and:
•
if
Level +
is selected, the triggering condition is fulfilled only when
Source
has the greater value than
determined by
Level
, otherwise the triggering condition is not fulfilled.
•
if
Level
–
is selected, the triggering condition is fulfilled only when
Source
has the lower value than this
determined by
Level
, otherwise the triggering condition is not fulfilled.
•
if
Slope +
is selected, the triggering condition is fulfilled only when the rising value of
Source
is passing
the level determined by
Level
.
•
if
Slope
–
is selected, the triggering condition is fulfilled only when the falling value of
Source
is passing
the level determined by
Level
.
•
if
Gr
is selected, the triggering condition is fulfilled only when the signal has the greater level
than determined by
Level
and the gradient of the signal is greater than determined by
Gradient
.
Otherwise the triggering condition is not fulfilled.
•
if
Gradient -
is selected, the triggering condition is fulfilled only when the signal has the lower level than
this determined by
Level
and the gradient of the signal is lower than determined by
Gradient
. Otherwise
the triggering condition is not fulfilled.
Step for checking the triggering
condition
The
Trig. Step
position enables the user
to select interval (integration period) for
condition
evaluation:
equal
to
Logger step
(
path:
<Menu>
/
Measurement
/
Data Logging
/
Logger Setup
),
100ms
,
1.0s
, and equal to
Meas. Period
(
path:
<Menu>
/
Measurements / General Settings
). If
Meas. Time
is selected the triggering
condition is checked every second and
RMS is averaged from the beginning of
the
measurement
(
Meas. Time
is
displayed in the right upper corner of the
display right under the Real Time Clock).
<
>
Source for triggering condition and
channel of triggering signal
The
Source
position enables the user to
select the type of source for triggering
condition calculation:
Vector
,
Profile 1
or
Spectrum
(available for
1/1 Octave
and
1/3 Octave
functions).
The
Channel
position enables the user to
select the channel of triggering source.
Depending on value of the
Source
parameter, the value of
Channel
will be
different.
<
>