![Svantek SV 106A Скачать руководство пользователя страница 69](http://html1.mh-extra.com/html/svantek/sv-106a/sv-106a_user-manual_1409261069.webp)
69
SV106A User Manual
8. SETTINGS OF THE INSTRUMENT PARAMETERS
– Instrument
The
Instrument
list contains different sub-lists and positions, which are
related to the settings of the hardware components of the instrument. In
order to open the
Instrument
list, the user has to press the
<Menu>
push-
button, select the
Instrument
position
and press
<ENTER>
.
The
Instrument
list content depends on
Instrument Mode
selection
from
the
Auxiliary Setup
menu:
Simple Mode
or
Advanced Mode
.
Advanced Mode
gives the user more options.
The
Instrument
list is used for setting the various parameters which are primarily dedicated to control of the
instrument hardware and contains the following positions:
Keyboard
enables the user to set the operating mode of the
<Shift>
,
<Alt>
and
<Start/Stop>
push-buttons.
Multifunction I/O
enables the user to select the available functionality of the I/O port.
Power Supply
enables the user to
check the power source of the instrument.
RTC
enables the user to set the Real-Time Clock.
Remote Control
enables the user to activate or deactivate error confirmation function.
Unit Label
enables the user to check the type of the instrument, its serial number and the
current software version installed and the standards the instrument fulfils.
8.1 Selection of keyboard modes
– Keyboard
The
Keyboard
sub-list enables the user to
programme the operation mode of the
<Shift>
,
<Alt>
and
<Start/Stop>
push-
buttons and options for instrument wakeup
upon pressing key when it works in the
screen saving mode.
<ENT>
<Shift> / <Alt> push-button mode
The
Shift Mode
position enables the user to choose the way of operation of
the
<Shift>
and
<Alt>
push-buttons with other ones. When the
Direct
option
is selected, the
<Shift>
and
<Alt>
push-buttons operate as in the keyboard of
a computer
– to achieve the desired result, the second push-button has to be
pressed at the same time as with
<Shift>
/
<Alt>
. When the
2nd Fun.
option is
selected the
<Shift>
/
<Alt>
push-buttons operate in the sequence with the
other one. This enables the user to use only one hand to operate the
instrument.