![Svantek SV 106A Скачать руководство пользователя страница 84](http://html1.mh-extra.com/html/svantek/sv-106a/sv-106a_user-manual_1409261084.webp)
84
SV106A User Manual
11. DOSIMETER FUNCTION
Activation of the dose meter function
The
Dosimeter
position activates or
deactivates the dose meter function. If the
Dosimeter
function is switched on, then all
channels will be assigned for whole-body
(
WBV
)
or
hand-arm
(
HAV
)
dose
measurement.
<ENT>
Dosimeter parameters can be setup in the
HAV/WBV Dosimeter
window, opened from the
Measurement
list.
If the
Dosimeter
function is switched on, then the
HAV/WBV Dosimeter
position in the
Measurement
list will
became active and
DLM
,
D1/1
or
D1/3
function abbreviations appear in the upper line.
11.1. Setting parameters for dose measurements
– HAV/WBV Dosimeter
The
HAV/WBV Dosimeter
list is opened from the
Measurement
menu.
This list enables the user to set up the parameters for human vibration dose
measurements, like: exposure period, type of measurement (whole body or
hand arm), performed in channels 1-3 and 4-6, action limits used for some
standards (
U.K.
,
Italy
,
Poland
,
France
,
Germany
,
China
,
Brazil
), as well
as limits, defined by the user (
User
).
Setting the measurement type for channels 1-3 and 4-6
Positions
1-3 Dosimeter
and
4-6 Dosimeter
enable the user to set the
desired type of the measurement, performed with the use of channels 1,2,3
and 4,5,6
– hand-arm (
HAV
) or whole-body (
WBV
) vibration.
Setting the exposure time
The
Exposure Time
position enables the user to set the desired value of
the exposure time that is used for the calculation of the
HAV/WBV Dose
results. The
Exposure Time
values are within the range [
00h01
,
24h00
].
Setting the standard for dose measurements
The
Standard
position enables the user to set the standards for the
measurements of the
HAV/WBV Dosimeter
. The available values of this
position are
U.K.
,
Italy
,
Poland
,
France
,
Germany
,
China
,
Brazil
and
User
.
Depending on the settings in the position
Standard
it is possible to view
(
U.K.
,
Italy
,
Poland
,
France
,
Germany
,
China
,
Brazil
) or edit (
User
) limits
for dose calculation.