33
SV106A User Manual
Activation of the measure trigger
The
Measure Trigger
position activates
or deactivates the measure trigger
function. This position doesn’t appear in
the
Simple Mode
. If the
Measure Trigger
function is switched off, then the
Measure Trigger
position
in
the
Measurement
list will be not active.
<ENT>
Activation of the dose meter function
The
Dosimeter
position activates or
deactivates the dose meter function. If the
Dosimeter
function is switched on, then
all channels will be automatically assigned
for the whole-body (
WBV
) or hand-arm
(
HAV
) dose measurement for ease of
operation by the user.
<ENT>
Dosimeter parameters can be setup in the
HAV/WBV Dosimeter
window, opened from the
Measurement
list. If the
Dosimeter
function is switched on, then the
HAV/WBV Dosimeter
position in the
Measurement
list will be active and
DLM
,
D1/1
or
D1/3
function abbreviations appear in the icon’s line of the display.
Setting the Logger mode
The
Logging Mode
position enables the user to deactivate the logger
function (
Off
) or to activate this function by choosing the logger mode
(
Logger
or
Wave
). The
Wave
option doesn’t appear in the
Simple Mode
(
path: <Menu> / Auxiliary Setup / Instrument Mode
). In case the
Logger
is
selected the time history of selected results will be saved in the logger file.
In case the
Wave
option is selected the time wave signals for the channels,
selected in the window
Wave Channels
, will be recorded in the logger file.
The logger file name is defined in the
Logger Setup
window and will have
predefined name - for the history results
&LOG#
and
&REC#
for wave
recording.
Logging Mode
affects the view of
Measurement
and
Data Logging
lists.
<ENT>
<ENT>
Data Logging screen view when Logger mode is switched on
<ENT>
<ENT>
Data Logging screen view when Wave mode is switched on