28
SV106A User Manual
Note: During the calibration measurement, the external disturbances (vibrations or acoustic noise)
should not exceed a value of 100 dB.
7. When calibration measurement is finished select
Accept
and press
<ENTER>
to store the calibration measurement result.
Select
Check
to run sample measurement simulating usage of the newly
calculated calibration factor, assuming that still the same excitation level is
applied to the transducer. In case calibration factor is correct, the measured
level should be in proper correspondence with calibrator level.
Select
Reject
to cancel calibration procedure.
4.2.6.
History of the calibrations
– Calibration History
The
Calibration History
window
displays up to ten last calibration
records.
<ENT>
To review the calibration record, the user has to use the
<
>
,
<
>
push-
buttons. The opened window will contain the date and time of the performed
calibration measurement, the way the calibration was done (
Calibr.
By Measurement
or
Calibr. By Sensitivity
or read from TEDS) and the
calibration factor (
Calibration Factor
) that was obtained.
If calibration measurements were not performed the
Calibration History
window does not contain any records. The content of this window is cleared
after the
Clear Calibr. History
command.
4.2.7.
Clear calibration records - Clear Calibr. History
The user can clear all stored calibrations records. To do this the user has
to choose the
Clear Calibr. History
position in the
Calibration
sub-list and
press
<ENTER>
to perform this command.
The instrument requests confirmation of the command. The next pressing
of the
<ENTER>
push-button, when the
No
option is selected, closes the
window and returns the instrument to the
Calibration
sub-list. If
Yes
is
selected the pressing
<ENTER>
will delete the history and return the
instrument to the previous menu.
After
Clear Calibr. History
command has been performed the
Calibration History
window does not any more contain records.