![Svantek SV 106A Скачать руководство пользователя страница 37](http://html1.mh-extra.com/html/svantek/sv-106a/sv-106a_user-manual_1409261037.webp)
37
SV106A User Manual
If second profile is active during
Dosimeter
measurements, the filter can be
HP
,
Vel3
or one of:
BL Wd
,
BL Wk
or
BL Wh
depending which filter was predefined for the first profile of the chosen channel, per the rule
described above.
RMS detector
The only one
1.0s
RMS detectors is available in the instrument.
5.4 Setting the vector parameters
– Vectors Setup
The
Vectors Setup
position enables
the user to select the coefficients to
calculate the vector for channels 1, 2, 3
and 4, 5, 6.
Vector is calculated based on different
set of coefficients for three axes (X, Y,
Z), which may be selected in the
Mode
position: for hand-arm measurements
(
Standard H-A
),
for
whole
body
measurements (
Standard WBV
), for
measurements
with
user
defined
coefficients (
User
), for MTVV (
MTVV
)
and for PPV (
PPV
) measurements. For
Standard H-A
and
Standard WBV
modes coefficients are predefined. For
the
User
,
MTVV
and
PPV
modes it is
possible to define coefficients for vector
calculation.
<ENT>
<
>
When the user needs to calculate a vector with other than standard
coefficients, it is possible to select the coefficient within the values from
0.00
to
2.00
.
The values presented above are taken into account during the
calculations of the measurement results.
VECTOR
is calculated
according to the formulae:
2
3
2
3
2
2
2
2
2
1
2
1
x
k
x
k
x
k
VECTOR
Where k
1
, k
2
and k
3
are the coefficients and x
1
, x
2
and x
3
are RMS results for different channels. It is important
that the user should choose only coefficients corresponding with the proper channels.
5.5 Setting the data logging functionality
– Data Logging
The
Data
Logging
list enables the user to program the logger functions:
recording of the results history, event and signal (wave) recording for all six
channels.
The
Data
Logging
list content depends on
Instrument Mode
(
path:
<Menu> / Auxiliary Setup
):
Simple Mode
or
Advanced Mode
. Some
additional functions like triggering, markers, event and wave recording
appear only in
Advanced Mode
.