40
SV106A User Manual
The view of the
Logger Results
list depends
on the settings of the
Enabled Channels
and
Enabled 2nd profiles
parameters
(
path:
<Menu> / Measurement / Channels Setup
).
In
Advanced Mode,
it is possible to define
logger results for each channel and profile
individually while in
Simple Mode
only for
group of channels
1-3
and
4-6
.
Selection results for channels and profiles
Depending on
Instrument Mode
and the
Enable 2nd profiles
parameter
the user may activate the results for channels and profiles
(
PEAK
,
P-P
,
MAX
,
RMS
and
VDV
), which will be recorded to the logger file (column
Log
), activate plot (column
Plot
) and select its colour (column
Color
) in the
windows with names:
Channels x-y / Channels x-y Profile z / Channel x
/ Channel x Profile y
.
The
VDV
history will not be recorded if
Wh
filer is chosen for this profile.
Activation / deactivation can be done by means of the
<
>
,
<
>
push-
buttons pressed together with
<Shift>
. The position is changing by means
of the
<
>
,
<
>
and
<
>
,
<
>
push-buttons.
The example display of Logger with two
selected results is attached.
=>
Selection vectors for logging
The
Vector
list enables the user to
activate the vectors (
VEC13
and
VEC46
), which will be recorded to the
logger file, activate plot and select its
colour.
<ENT>