SV106A User Manual
<=>
<=>
9.4. Reference signal in vibration measurements - Reference Levels
The
Reference Levels
sub-list enables the
user to set the reference level of the
vibration signal. The values, which are set
here,
are
considered
during
the
calculations of the measurement results
expressed in the
Logarithmic
scale (with
the
dB
as the units).
<ENT>
In the
Acceleration
position the user can set the reference level of the acceleration signal from
1
µms
-2
to
100
µms
-2
. In the
Velocity
position the user can set the reference level of the velocity signal. It is possible to
set this level from
1 nms
-1
to
100 nms
-1
.
9.5. User filter setting
– User Filters
The
User Filters
sub-list enables the user to introduce the values of the
coefficients of the user defined frequency filters. This position is active only
in
1/1 Octave
and
1/3 Octave
modes. This sub-list is described in Chapter
10.
9.6. Selection of the Vibration units - Vibration Units
The
Vibration Units
sub-list enables the
user to select the units for the Vibration
measurements.
It is possible to select the
Non-Metric
units
(e.g. g, ips, mil etc.) or
Metric
units (e.g.
m/s
2
, m/s, m etc.).
<ENT>