5.2.5
Range test
The range test is designed to be able to test the range of the alarm transmitter to ensure that it has range over the
area you have in mind, such as the entire area of the care recipient's apartment. The range test also provides
information as to the battery level of the alarm transmitter.
Press the green button and hold.
Also press the red and grey buttons and hold them down until an ascending beep is heard. Then release all
buttons.
Now you are inside the programming mode, as indicated by the light switching between the LEDs.
Press the grey button down until you hear three consecutive beeps.
Release the button and the carephone then replies with four beeps.
With maximum 1 minute intervals, the alarm transmitter can now be tested.
o
One short beep means that the accessory is programmed and the battery is OK.
o
One long beep means that the accessory is programmed, but the battery level in the accessory
is low and needs replacing.
o
Two short beeps means that the accessory being used has not been programmed for the
current carephone but the alarm gets through and the battery is OK.
o
Two long beeps means that the accessory being used has not been programmed for the
current carephone, and the battery level in the accessory is low and needs replacing.
Remember to test the range everywhere in the apartment as there may be materials in walls, etc. that
obstruct the radio signal from the alarm transmitter.
When one minute has passed without a test, the range test is automatically terminated. To end the range
test manually, just press the green button.
Then press the green button again to exit the programming mode.
Note! You can conduct the range test as per the previous carephones, by turning off the carephone and holding
the green button when you restart the carephone.
6
Battery Replacement
6.1
Battery replacement for Careline Anna/GSM
When the battery needs replacing (i.e., when the carephone sends accumulator alarms) do the following:
Remove the back cover of the carephone.
Switch off the appliance using the power switch and unplug all connectors from the carephone.
Remove the screw from the battery holder (see image 9) with a T8 torx screwdriver. Pull out the battery
holder (see image 10).
Disconnect the cable on the spent battery and connect a new battery (see image 11).
Slide the battery holder in place (press the SIM card if the battery holder does not slide easily) and attach it
with the screw
Replace all cables and the cover.
Image 9. Unscrew the battery holder. Image 10. Pull out the battery holder. Image 11. Loosen the cable and
connect the new battery.