Programming and installation manual Careline Anna / Careline GSM
Doc. no. 32-022-03-01 EN v6
4.2.5
Programming of Sequences
The programming of sequences defines which alarm types shall call which alarm receiver and in what order.
(See table in Appendix A.) Follow the instructions below
Choose
Sequences
,
press
Yes
Choose
Def. sequence
, press
Yes
Choose
Sequence 1
, press
Yes
Preset value
Choose new value
Use
to switch between the eight values and positions and
to change the value. The letters
represent the listed phone numbers and addresses. If no telephone number is required for the current
position, enter "-".
When settings are entered, press
Yes
.
Repeat this procedure for the other sequences that you need.
Press
No
to return to the "Sequencing" menu.
Choose
Alarm type
,
press
Yes
Choose
Medical alarm
,
press
Yes
Choose
Sequence 1,
press
Yes
Choose which sequence shall be used for the different alarm types. For example, you can set it so that all
care alarms ring as "Sequence 1 Press
and
until the right sequence is shown
New value
.
Press
Yes
to accept.
Repeat the procedure for all alarm types.
Press
No
to exit the menu "Types of alarm".
Choose
Repeat seq
., press
Yes
Value
1
has been preset
Delete the existing value with
and enter the desired number of sequence repetitions and press
Yes
.
Press
No
once to leave the Sequences menu.
Location code
Alarm codes
Protocol
>Sequence
s
<
SEQUENCES
>
Def. sequence
<
Def. alarmtypes
Silent/Speech
DEF. SEQUENCE
>
Sequence 1
<
Sequence 2
Sequence 3
Value in HT:
A B C D - - - - - -
New value:
A B C D - - - - - -
Value in HT:
A B C D - - - - - -
New value:
A B C D - - - - - -
SEQUENCES
Def. sequence
>
Alarm Types
<
Series connection
ALARM TYPES
>
Medical Alarm
<
Passive Alarm
Acc./Mains
Value in HT:
<Sequence 1>
New value:
<
Sequence
1
>
Value in HT:
1
New Value:
1
Alarm Types
Cascading
Silent/Speech
>
Repeat seq.
<