8
Stille imagiQ2
TM
service manual Overview
Ov
er
view
Start-Up procedure
Follow the procedure described below before the
Stille imagiQ2
TM
operation table is used for the first
time:
n
Before the table is used for the first time:
- charge the batteries until they are fully charged,
(the main switch must be “ON” when charging
the table.)
- Fill out the installation report.
- Fill out the testprotocol ISO -EN 62353
1)
Read the User Manual
2)
Test run all functions on the hand control as
described:
3)
Column, Run column from lowest position to
highest position and up again.
4)
ISO ROLL™/Lateral tilt right, Run tabletop from
zero position to maximum right inclination, and
then zero the table.
5)
ISO ROLL™/Lateral tilt left, Run tabletop from
zero position to maximum left inclination, and
then zero the table.
6)
Longitudinal tilt (Trendelenburg), Run tabletop
from zero position to maximum head down, and
then zero the table.
7)
Reversed longitudinal tilt (anti-Trendelenburg),
Run tabletop from zero position to maximum
head up and then zero the table.
8)
Quick-trend (Trendelenburg), Run tabletop
from zero position to maximum head down and
then zero the table.
9)
Zero position. Lateral tilt, Run tabletop from
maximum left inclination to maximum right
inclination and back. Make sure that the table
stops both times it passes zero position.
10)
Longitudinal tilt (Trendelenburg) Run tabletop
from maximum head down to maximum head
up and back. Make sure that the table stops
both times it passes zero position.
11)
True Free Float™ /
Panning. Pan functions,
place table in 0-position and test run both
axis. Check that motion works effortlessly and
smooth (manoeuvre with one hand only).
12)
All above mentioned steps shall be controlled
on the Safety Hand Control as well.
Daily control before procedures
n
The operator should have knowledge how to
operate the device, be aware of warnings and
signals described in the user manual and the
user manual should always be available.
n
The operator should perform a function Check
accordingt to 2-11.
n
Test the device for proper function and inspect
cables for cuts and other damage. If in doubt,
replace the relevant parts.