service manual Overview
Ov
er
view
Periodical preventive maintenance (PM)
The imagiQ2 requires a maintenance period of 1time every year or 2000 hours of service
whichever comes first according to the service schedule in this manual.
(2000 hours equals to 8 hours of service 5 days / week for 1 year)
If the service is not correctly documented the warranty is void.
Stille AB can not guarantee the product’s safety if Periodical preventive maintenance (PM) has
been neglected.
Cleaning and care of the polyurethane coated mattress
General Guidance
Attention must be paid to the properties of any other materials, (eg washing agents and
washing instructions).
Some surface wrinkling may result from cleaning procedures. This has no adverse effect on
the fabric’s properties.
Do not use washing machines
Abrasive cleaning agents should not be used.
Washing and Disinfection
Superficial dirt on the coating may be removed by wiping with a soft cloth moistened with
water containing a neutral detergent. More persistent contamination may be treated by wiping
with alcohols or turpentine substitute, followed by hot water and detergent.
Routine cleaning and disinfection may be carried out on the coating with hand hot water
and a neutral detergent or with a sodium hypochlorite solution (0.1% or 1000 parts per mil-
lion available chlorine).
The material is compatible with the 10,000ppm available chlorine in solution required for
the decontamination of blood spills. After the 2 minute required dwell, excess solution must be
removed and the surface thoroughly rinsed and dried prior to reuse or storage.
Proprietary disinfectants may be used provided manufacturer’s instructions are followed.
All cleaning agents, and disinfectants, must be thoroughly rinsed off and the item dried
before storage. Failure to do this may result in the accumulation of reagent that could damage
the polyurethane coating, react with the bed frame, or negate the biocompatibility results of
the fabric.
Drying
It is essential that articles be thoroughly dried after all cleaning procedures and before
storage.
Drying may be effected at temperatures up to 130 °C (266°F).
Storage
Store in a cool, dry area. Avoid excessive pressure and contact with non-colourfast materi-
als