Stille imagiQ2
TM
service manual Technical
Technic
al
TROUBLE SHOOTING GUIDE
Electrical problems, problems with the hand control
Symptom
Probable cause
Procedure
The hand control function but
show a different status.
a. Unserviceable hand control
b. Unserviceable emergency
control.
c. Unserviceable
control unit.
a. Replace the hand control.
b. Replace the emergency
control.
c. Replace the control unit.
Hand control appears to
function normally but the
hydraulic pump remains silent
and only return movements
are possible. e.g. Column
down, unparking etc.
a. Connector to pump motor
incorrectly connected.
b. Pump motor unserviceable.
c. Control unit unserviceable,
a. Check the connector.
b. Check the pump motor.
c. Replace the main board.