Stille imagiQ2
TM
service manual Safety Information
Stille imagiQ2
TM
service manual Safety Information
SAFETY INFORMATION
Always follow the manufacturer’s instructions!
Read carefully and always follow the instructions in the users manual before first using the imagiQ2™ operating
table. This manual has been written to ensure that you will handle the table in a safe and correct manner, use this
manual together with the users manual.
n
The device is intended to be used by qualified surgical staff.
n
Quick Stop/ Main switch: Move the switch downwards for immediate shut-off (see page 12).
n
All movements are interrupted when the hand control is switched off (OFF), or when the Quick Shut-off
switch is turned off.
n
All table top movements cease immediately if the button in question is released.
n
Charge the batteries if the battery indicator is indicating low battery.
n
Avoid discharging the batteries completely. Maximum battery life-time is achieved if charged regularly.
n
The manual must always be stored within reach of the operating table.
n
The main switch must be turned “ON” when charging the table.
n
Follow the packaging / unpackaging procedure to avoid personal injuries, equipment damage or property
damage.
n
This device, as all medical electric equipment in general, need special precautions regarding EMC and need to
be installed and put into service according to information in the user and service manual.
n
During operation make sure to not touch external ports or conenctors and patient at the same time.
n
Always use anti-static cushions to ensure that static electricity is discharged to the anti-static floor.
n
The table is intended to be used on an anti-static floor.
n
Always connect the table’s potential equalization connection point to protective earth.
n
Never place surgical cloths/ sheets so that the grounding rails are totally covered. Surgical cloths/ sheets pre-
vent electric discharge to anti-static floors.
n
Radio transmitting equipment, cellular phones etc. shall not be used in the close proximity of the imagiQ2™
table since this could influence the performance of the imagiQ2™ table.
n
Particular precaution must be considered during use of strong emission sources such as High Frequency
surgical equipment and similar so that e. g. the HF-cables are not routed on or near the imagiQ2™ table. If in
doubt, contact qualified technician or your local representative.
n
Do not subject the operating table to electromagnetic fields that exceed the applicable standards for radio
frequency equipment, such as diathermy apparatus, defibrillators and defibrillator monitors.
n
The table is only to be used with medical equipment complying with EN 60601-1, EN 60601-1-2.
n
The imagiQ2™ table complies with the EMC requirements according to IEC 60601-1-2.
n
The device is not intended for use with flammable anaesthetic gases. A possible explosion hazard exists and
personal injury or equipment damage could occur.
n
The device shall be cleaned / disinfected according to the user manual.
n
Do not immerse the control unit device in water or any other liquid (see the care / maintenance section for
specific details).
n
The operating table is disconnected from mains by disconnecting the AC-power cord from the table.
n
Unplug the power connector from its power source before moving the table, cleaning or servicing. Failure to
do so could result in equipment damage.
n
The table is more stable when parked. Always park the table when it’s not being transported.
n
Only use accessories approved by STILLE AB.
n
Always check that mounted accessories do not collide with the operating table or surrounding equipment
when movement is activated
n
Remove any obstacles before lowering or tilting the operating table. It could tip over.
n
The operating table base must always be stable and horizontally level.
n
Centre the table top longitudinally over the column as far as possible before lowering the free end of the table
top.
n
When adjusting or when panning the operating table, never put your fingers under the table top or cover. Risk
of injury.
n
Secure the patient before changing the position of the operating table.
n
The patient must not lie on conductive mattresses or on damp bedding when defibrillators, defibrillator moni-
toring screens or high frequency apparatus are being used. Nor should the patient come into contact with
metal parts or their accessories on the operating table. This could result in burns.
n
Keep the patient under constant observation.
n
Use only STILLE spare parts.