To replace Power supply
Warning take measure for ESD protection
1)
If possible place the table in position for service according to instruction, mini-
mum try to elevate the table top to the highest position, lock the wheels.
2)
Remove the emergency cover, the lowest column cover, and the front cover and
then the protective plates under the covers.
3)
Perform an electric LoTo according to the instruction to remove any stored
energy in the system.
4)
Protect the electrical components against ESD.
5)
The Power supply is mounted under the main board. Make a note of where the
electrical connectors are placed on the main board and carefully remove all connectors,
pull straight up, do not “wiggle” the connector, you might break it.
6)
The main board assembly is attached to the frame with 2 screws from the outside,
remove the screws and lift up the assembly.
7)
The main board is mounted with 6 screws, remove the screws and remove the
main board. NOTE do not overtighten the screws when attaching the main board.
8)
Replace the power supply.
9)
Reinstall the connectors and restore the power according to the LoTo procedure.
10)
Test the functions, take care as the device is not protected by the covers, risk for
personal injury or damage to the device.
11)
Reinstall the protective covers.
12)
Complete the LoTo procedure.