44
Stille imagiQ2
TM
service manual Technical
Technic
al
TROUBLE SHOOTING GUIDE
Electrical problems, Batteries
Symptom
Probable cause
Procedure
The table is ”completely
dead” when under battery
operation. The green ”Fully
charged” lamp on the hand
control illuminates after a
few hours charging but the
lamps extinguish (almost)
when an attempt is made to
run the table while charging
is in progress(and a slight
squeaking noise can be heard
from the hydraulic motor)
after which only the blue
”Charging in progress” lamp
illuminates when the button
is released.
a. The battery package is not
correctly connected in the
table.
b. The battery fues has
blown.
c. The battery package is
unserviceable/failure in a cell.
a. Check the connections of
the battery package
b. Check first all the heavy
red/black cables in the table
for damaged insulation/a
short circuit. Check the bat-
tery fuse.
c. Replace the battery pack-
age.
The batteries are dead, table
will not work, batteries will
not take charging
The ”Charging in progress”
lamp is not illuminated
a. Power cord is not con-
nected to a socket with AC
power
b. Check the fuse box for
blown fuses
c. Check the output from
the AC inlet, should be 100/
120/ 230V
d. Check 24V from Power
supply to main board should
be about 24V.
e. Check the output from the
charging circuit it should be
about 27-29,5VDC
a. Try changing to another
socket.
b. Change the fuses.
c. Change the ac inlet.
e. Replace thePower supply.
e. Replace the main board.