46
Stille imagiQ2
TM
service manual Technical
Technic
al
TROUBLE SHOOTING GUIDE
Hydraulic problems pump motor
Symptom
Probable cause
Procedure
Hydraulic pump unit
No movement of the table
is executed when requested
from the hand control.
a. When activating a move-
ment on the hand control the
hydraulic pump motor does
run and the solenoids become
activated.
b. When activating a move-
ment on the hand control the
hydraulic pump motor does
run and the solenoids become
activated.
c. When activating a move-
ment the solenoids do not
become activated but the
motor is running.
d. The hydraulic pump motor
does not run.
Check 24VDC to the hydrau-
lic motor.
a. Check the cable connec-
tions on the main board.
b. replace the hydraulic pump
motor.
c. Replace the main board
d. Replace the main board.