20
Stille imagiQ2
TM
service manual Technical
Technic
al
ENERGY CONTROL PROCEDURES
(OSHA 29 CFR 1910.147(d) Application of Control)
ELECTRICAL ENERGY:
1.
Notify all affected individuals of a shutdown.*
2.
Position Table for service
3.
Turn OFF main switch
4.
Remove A/C plug from wall power and from table.
5.
Apply the red lock and tag to the AC plug.
6.
Turn ON Main switch to verify primary power isolation.
7.
Remove covers and protective plates
8.
In the Table base verify zero volts A/C at AC-connector
9.
Disconnect the red/black battery connector
10.
Verify isolation by pressing ON on the hand control, nothing shall happend.
11.
Turn OFF main switch
11.
Electrical power is now locked out.
12.
It might be necessary to do an hydraulic power LOTO according to the procedure for
”hydraulic energy”.
*In the event service will continue into another day, and/or the machine will cause a hazard if
reenergized after the authorized employee leaves the location, leave the A/C cord cover installed
with a “transition” lock and yellow tag attached. (OSHA 29 CFR 1910.147(f)(4) Shift/Personnel
Changes)
*If more than one employee is going to conduct service on the same system, at the same time,
multiple locking devices must be employed an each authorized employee must adhere the entire
LOTO procedure. (OSHA 29 CFR 1910.147(f)(3) Group Lockout/Tagout)
TESTING AND POSITIONING DURING SERVICE
(OSHA 29 CFR 1910.147(f)(1) Testing/Positioning)
In the event that lockout/tagout devices must be temporarily removed from the energy isolating
devices, for testing or positioning purposes, the following sequence shall be followed:
1.
Clear the table of tools and materials.
2.
Ensure that all employees/individuals in area are safely positioned or removed.
3.
Remove energy control measures.
4.
Energize system and proceed with testing or positioning.
5.
De-energize table again after testing or positioning.
RESTORE POWER AFTER SERVICE:
(OSHA 29 CFR 1910.147(e) Release of Lockout/Tagout)
1.
Re-inspect work area for hazards.
2.
Reconnect the red/black battery connector
3.
Reinstall table covers.
4.
Turn on main switch.
5.
Notify all affected personnel that power being restored.
6.
Remove red lock and tag
7.
Energize the table.
8.
Verify proper system operation.
9.
Inform customer that system is ready to be returned to service.