14
Stille imagiQ2
TM
service manual Technical
Technic
al
Position the table for service and repair.
To position the table for service (and preventive maintenance) time ~20 minutes.
1) Make sure the table is fully charged(reccomendation)
2) Move the table column to the highest position
3) Park the table
4) Remove the emgergency cover
5) Remove the lower coumn cover
6) Remove the front cover( be careful with the cables when the column covers is slid out
7) remove the front plate connection plate
8) Remove the lateral side plates
9) Remove the table top
10) remove the top cover
11) Depending on the service or repair additional dismantling can be necessary, see separate
instruction.