service manual Technical
Technic
al
TROUBLE SHOOTING GUIDE
Electrical problems, indications on the hand control unit
Symptom
Probable cause
Procedure
The green ”Fully charged”
lamp on the hand control
illuminates after a few hours
charging, despite the ”Low
battery level” lamp flas-
hing when the charging was
begun.
a. The battery package is
unserviceable.
b. The battery charger is
unserviceable.
c. The main board/ control
unit is unserviceable.
a. Replace the battery pack-
age.
b. Replace the main board.
c. Replace the main board.
The green ”Fully charged”
lamp on the hand control
does not illuminate after
more than 10 hours charging
with the yellow ”Charging in
progress” lamp illuminated.
a. The battery package is
unserviceable/ short circuit in
a cell.
b. The battery charger is
unserviceable.
c. The control unit is unservi-
ceable.
a. Replace the battery pack-
age.
b. Replace the main board.
c. Replace the main board.