service manual Technical
Technic
al
To replace Inclinometer board
Warning take measure for ESD protection
1)
If possible place the table in position for service according to instruction, or
try to elevate the table top to the highest position, lock the wheels.
2)
Remove the table top and the top cover, the emergency cover and then the
protective plates under the cover.
3)
Perform an electric LoTo according to the instruction to remove any stored
energy in the system.
4)
Protect the electrical components against ESD.
5)
The inclinometer board is mounted with 4 screws, remove the screws.
6)
Carefully remove the connector, pull straight up, do not “wiggle” the con-
nector, you might break it.
7)
Reinstall the card and connector and mount it, restore the power according
to the LoTo procedure. NOTE! do not overtighten the scews tighten only lightly.
8)
Test the functions, take care as the device is not protected by the covers, risk
for personal injury or damage to the device.
9) To calibrate a new 0-position, place the table in horizontal position with the
table top slightly raised in the front about 1-2 degrees up, turn OFF the hand
control and press the 0- button on the main board for 1 second, tunr table on and
check 0 position, if needed redo the previous steps.
10)
Reinstall the protective covers.
11)
Complete the LoTo procedure.