22
Stille imagiQ2
TM
service manual Technical
Technic
al
To replace Connection board
The connection board is equipped with connectors for:
2 pan handle connectors (2 pan handles can be connected at the same time)
2 hand control connectors (only 1 hand control can be connected at the same time)
1 Accessory control connector ( accessory table tops under development)
1) If possible try to elevate the
table top to the highest
position, lock the wheels.
2) Remove the lateral cover over
Connection board, then
remove the emergency cover
and then the protective plate
under the cover.
3) Perform an electric LoTo
according to the instruction to
remove any stored energy in
the system.
4) Protect the electrical
components against ESD.
5) Make a note of where the electrical con-
nectors are placed and carefully remove them,
pull straight up, do not “wiggle” the connector,
you might break it.
8) Remove the 4 screws that holds the connec-
tor board bracket.
7) The Connection board is mounted with 5
plastic nuts, use a small screw driver to unlock
the “lock lip” on top of the nut, remove the
nuts and replace the main board.
8) Reinstall the connectors and restore the
power according to the LoTo procedure.
9) Test the functions, take care as the device is
not protected by the covers, risk for personal
injury or damage to the device.
10) Reinstall the protective covers.
11) Complete the LoTo procedure.