28
Stille imagiQ2
TM
service manual Technical
Technic
al
To replace Lateral cylinder
1)
If possible place the table in position for service according to instruction,
minimum try to elevate the table top to the highest position, lock the wheels and
0-out the table.
2)
Remove the table top, all the top covers, the emergency cover the lowest
column cover and the front cover and then the protective plate under the cover
emergency cover.
3)
Perform a hydraulic and electric LoTo according to the instruction to
remove any stored energy in the system.
4)
Place some paper towels under the hoses and then carefully untighten the
hoses, some amount of oil will come from the hoses, wait until it stops (a couple of
seconds) if oil with high pressure is flowing from the hoses, check that the LoTo is
correct performed.
5)
The cylinder is attached to the frame with 2 circlips, remove the circlips and
remove the cylinder.
6)
Mount the new cylinder.
7)
All washers must be replaced and tighten with the correct torque (12Nm)
all other connectors shall be tightened finger tight and then minimum 3/4 turns.
8)
Restore the power according to the LoTo procedure.
9)
Inspect and refill the oil level
10)
Reinstall all parts.