13
Stille imagiQ2
TM
service manual Overview
Ov
er
view
Installation of the imagiQ2
Complete the installation report imagiQ2
Document number 0260-6PF002 and submitt
to [email protected]
Planning reference data
Transport data:
Length =2400 mm / 7,8 ft
Width= 875 mm / 2,9 ft
Height = 1168 mm / 3,8 ft
Gross weight 450 kg / 992 Lbs
Net weight 265 kg / 584 Lbs
Transport of Dangerous Goods by Air – Limi-
ted to class/UN 3164 –ADR, 1.3/IMDG 1.3/
IATA-DGR 1.5
Electrical data:
Mains voltage 100 - 230 V AC,
50 / 60 Hz
Rated power Max 504 VA
Fuse rating
T6,3 H, 250V SB
Output power 400W
Environmental data:
Operating conditions
Temperature: +10°C to +40°C (50°F
to104°F)
Humidity: 30% - 75%.
Atmospheric pressure 700hPa - 1060hPa
Storage conditions /Transportation
requirements
Temperature -40°C to +70°C (-40°F
to158°F)
Humidity 10% - 80%rh. incl. condensation.
Atmospheric pressure 700hPa - 1060hPa.
Tools for installation, service and repairs
Knife or scissor
Screw drivers
Small = 2 mm blade width
Medium = 5 mm blade width
Large = 10 mm blade width
Screwdriver for cross head screws
Small = PH 0
Medium = PH 2
Torx
TX 10
Allen keys
1,5 mm 5 mm
2 mm 6 mm
2,5 mm 8 mm
3 mm 10 mm
4 mm
Combination wrenches
6 mm 13 mm
7 mm 14 mm
8 mm 15mm
10 mm 17 mm
11 mm 19 mm
12 mm
Other tools
Multimeter
Torque wrenche
A digital angle-measuring level (recommen-
ded least resolution = 0,1°) analog can also
be used
Hammer
Locking ring pliers (opening)
Other working materials:
Cable straps, 3 mm and 5 mm width.
Thread-locking fluid (Loctite 242 or equal)
Thread-seale (Scantech flange and thread
seale, 4000T, or Loctite Thread Sealant
542 Fine Threads)
AFA grease, Part number 507-K9038
Hydraulic Oil, part number 507-K90385
Содержание 530-1700