To replace Lift cylinder
1)
If possible place the table in position for service according to instruction,
minimum try to elevate the table top to the highest position, lock the wheels.
2)
Remove the table top, all the top covers, emergency cover the column
covers and the front cover and then the protective plate under the emergency cover.
3)
Float the table top to the left or right hand side, if that is not possible, tilt the
lateral tilt to full inclination either left or right side.
4)
Perform a hydraulic and electric LoTo according to the instruction to
remove any stored energy in the system.
5)
Place some paper towels under the hoses and then carefully untighten the
hoses, some amount of oil will come from the hoses, wait until it stops (a couple of
seconds) if oil with high pressure is flowing from the hoses, check that the LoTo is
correct performed.
6)
The cylinder is attached to the frame with 2 bolts, remove the bolts and
remove the cylinder up through the construction.
7)
Mount the new cylinder.
8)
All connectors shall be tightened finger tight and then minimum 3/4 turns.
9)
Restore the power according to the LoTo procedure.
10)
Inspect and refill the oil level
11)
Reinstall all parts.