TROUBLE SHOOTING GUIDE
True Free Float™ function
Symptom
Probable cause
Procedure
The free float does not work
properly.
a. The clamp locks (brakes)
does not open.
b. The hydraulic pump does
run, but no function is avai-
lable.
c. The hydraulic pump does
run constantly does not shut
off, but builds no pressure
a. Check the pressure switch.
b. Check that the solenoids
gets signal
c. Check the pressure switch,
try to adjust or replace the
switch.
Free float not possible in
longitudinal. Lateral works
fine.
a. The table is not in 0-posi-
tion
b. 0-position is not working
a. Press and hold 0 button on
the hand control to reset the
table to 0-position
b. Check the Inclinometer
card.
Free float not possible
a. Pan handle broken
b. The connection board is
damaged
a. measure pan handle
b. try to use the other pan
handel connector on the main
board, replace the connection
board.