PAT-80
● PAT-85 ● PAT-86 – USER MANUAL
6
2
General description and features of the instrument
PAT-80/85/86 digital tester is intended to measure the basic parameters of portable electrical ap-
pliances (power tools, household appliances, etc.) important for their safety: protective conductor re-
sistance, insulation resistance, continuity of connections, leakage current and RCDs.
The tester may be used to test equipment in accordance with the following standards:
EN 60745-1 Hand-held motor-operated electric tools. Safety. Part 1: General requirements.
EN 61029 Safety of transportable motor operated electric tools. General requirements.
EN 60335-1 Household and similar electrical appliances. Safety. Part 1: General requirements.
EN 60950 Information technology equipment
– Safety- Part 1: General requirements.
EN 61557-6 Electrical safety in low voltage distribution systems up to 1000 V AC and 1500 V DC
- Equipment for testing, measuring or monitoring of protective measures - Part 6: Effectiveness of
residual current devices (RCD) in TT, TN and IT systems.
VDE 0404-1 Testing and Measuring Equipment for Checking the Electric Safety of Electric De-
vices - Part 1: General Requirements.
VDE 0404-2 Testing and Measuring Equipment for Checking the Electric Safety of Electric De-
vices - Part 2: Testing Equipment for Tests after Repair, Change or in Case of Repeat Tests.
VDE 0701-0702 Inspection after Repair, Modification of Electrical Appliances.
Repeat Testing of Electrical Equipment. General Requirements for Electrical Safety.
AS/NZS 3760:2010 In-service safety inspection and testing of electrical equipment.
Basic functions of the instrument
Measurement of network voltage and frequency
Measurement of protective conductor resistance (protection class - I)
technical measurement method
measurement with sinusoidal current of network frequency and following currents:
200 mA
10 and 25 A
adjustable measurement time
adjustable upper limit in the range of: 10
mΩ…1.99 Ω with 0.01 Ω resolution
Measurement of insulation resistance
Four test voltages:
100 V
250 V, 500 V
1000 V
measurement of insulation resistance up to 599 M
automatic discharge of the capacitance of tested object after the insulation resistance meas-
urement is completed
adjustable measurement time
adjustable lower limit within the range of 0.1
MΩ….9.9 MΩ with 0.1 MΩ resolution
Measurement of substitute leakage current
adjustable measurement time
adjustable upper limit in the range of: 0.01 mA ... 9.9 mA with 0.01 mA / 0.1 mA resolution
Measurement of PE leakage current
adjustable measurement time
adjustable upper limit in the range of: 0.01 mA ... 9.9 mA with 0.01 mA / 0.1 mA resolution
measurement of the current with clamp