PAT-80
● PAT-85 ● PAT-86 – USER MANUAL
43
4.5
Measurement of leakage current I
PE
NOTE!
For Class I appliances, the measurement only makes sense if the R
PE
measurement
was positive.
Press
I
PE
.
Before starting the measurement, similarly to
sec. 4.2
, following parameters must be set:
test dura-
tion, limit, polarity
and
method
.
In
Change polarity
select:
yes
if the measurement is to be
repeated for reverse polarity,
no
if the measurement is per-
formed for only one polarity.
In
Methods
select:
socket
if test socket is to be
used for the measurement, and
go to step
,
clamp
if clamp is to be used for
the measurement, and go to step
,
PRCD
if you want to test a de-
vice with built-in RCD, and go to
step
.
Measurement with test socket
Connect the mains plug of the tested appliance into the test socket of the tester
. Addi-
tionally, it is possible to carry out the measurement with the probe connected to
T1
terminal
socket
.