PAT-80
● PAT-85 ● PAT-86 – USER MANUAL
83
13.2 Measurement of PE resistance
Measurement of resistance of protective conductor I = 200 mA (only Protection Class I)
Range
Resolution
Basic uncertainty
0.00
Ω...0.99 Ω
0.01
Ω
(4% m.v. + 2 digits)
1.00
kΩ...19.99 Ω
(4% m.v. + 3 digits)
Influencing factor
Designation
Additional uncertainty
Position
E
1
0%
Supply voltage
E
2
0%
Temperature
E
3
0.1% / ºC for R ≥ 0.5 Ω
0% / ºC for R < 0.5 Ω
unloaded output voltage: 4 V... 12 V AC
Test current: ≥200 mA for R = 0.2 Ω... 1.99 Ω
adjustable upper limit in the range of: 10 mΩ …1.99 Ω with resolution 0.01 Ω
adjustable measuring time: 3
s…180 s with a resolution of 1 s and option
Continuous test
Measurement of resistance of protective conductor I = 10 A (only Protection Class I)
Range
Resolution
Basic uncertainty
0 m
Ω...999 mΩ
1 m
Ω
(3% m.v. + 4 digits)
1.00
Ω...1.99 Ω
0.01
Ω
Influencing factor
Designation
Additional uncertainty
Position
E
1
0%
Supply voltage
E
2
0%
Temperature
E
3
0.1% / ºC
unloaded output voltage: <12 V AC
test current: ≥10 A for R ≤ 0.5 Ω
adjustable upper limit in the range of: 10 mΩ …1.99 Ω with resolution 0.01 Ω
adjustable measuring time in the range of:
3 s…180 s with resolution of 1 s
Measurement of resistance of protective conductor I = 25 A (only Protection Class I)
Range
Resolution
Basic uncertainty
0 m
Ω...999 mΩ
1 m
Ω
(3% m.v. + 4 digits)
1.00
Ω...1.99 Ω
0.01
Ω
Influencing factor
Designation
Additional uncertainty
Position
E
1
0%
Supply voltage
E
2
0%
Temperature
E
3
0.1% / ºC
unloaded output voltage: <12 V AC
test current: ≥ 25 A for R ≤ 0.2 Ω
adjustable upper limit in the range of: 10 mΩ …1.99 Ω with resolution 0.01 Ω
adjustable m
easuring time in the range of: 3 s…180 s with resolution of 1 s