PAT-80
● PAT-85 ● PAT-86 – USER MANUAL
42
For Class II and accessible parts disconnected from PE in Class I: connect the probe to
T2
terminal socket and touch the accessible parts of the tested appliance.
Press
START
.
Upon completing the measurement read the result. You can finish the test
before the defined test time duration by pressing
STOP
.
Positive test result: I
SUB
≤ LIMIT
Negative test result: I
SUB
> LIMIT
Tested appliance must be turned on.
Test circuit is electrically isolated from the mains and from mains' PE lead.
Test voltage is 25 V...50 V RMS.