Measurement of welding machines
In the test selection menu press
Welding machines
to go to a submenu with welding machine tests.
Tests not described in this chapter are carried out identically to the corresponding manual tests, pre-
sented in
All welding measurements
can also be made with weld-
ing cables
– but
only if the
cables are functional
.
5.1
Measurement of R
ISO
(LN-S, PE-S) in welding machines
Tests consist of measuring insulation resistance:
between shorted primary side conductors (L and N) and secondary winding of the welding ma-
chine (R
ISO
LN-S),
between the PE conductor and the secondary winding of the welding machine (R
ISO
PE-S).
NOTE!
For Class I appliances, the measurement only makes sense if:
the R
PE
measurement was positive and
the standard R
ISO
measurement was positive.
Press:
R
ISO
LN-S
. Depending on
the tested appliance go to
step
or
.
R
ISO
PE-S
. Depending on
the tested appliance go to
step
or
.
Before starting the measurement, similarly to
sec. 4.2
, following parameters must be set:
test volt-
age U
ISO
,
test duration
and
limit
.
Tested appliance must be turned on.
Test circuit is electrically isolated from the mains and from mains' PE lead.
Test result should be read only after displayed values are stabilized.