4.6
Measurement of differential leakage current I
Δ
Press
I
Δ
.
Before starting the measurement, similarly to
sec. 4.2
, following parameters must be set:
test dura-
tion, limit, polarity
and
method
.
In
Change polarity
select:
yes
if the measurement is to
be repeated for reverse polari-
ty,
no
if the measurement is per-
formed for only one polarity.
In
Method
select:
socket
if test socket is to be
used for the measurement,
and go to step
,
clamp
if clamp is to be used
for the measurement, and go
to step
,
PRCD
if you want to test a
device with built-in RCD, and
go to step
.
Measurement with test socket
Connect the mains plug of the tested appliance into the test socket of the tester.