PAT-80
● PAT-85 ● PAT-86 – USER MANUAL
38
4.3
Measurement of insulation resistance R
ISO
NOTE!
For Class I appliances, the measurement only makes sense if the R
PE
measurement
was positive.
Press
R
ISO
.
Before starting the measurement, similarly to
sec.
, following parameters must be set:
test volt-
age
,
test duration
,
limit
and test method:
probe-socket
(measurement between shorted L-N and
PE of the test socket or T2 probe),
probe-probe
(measurement between T1 and T2 probes) or
IEC
(IEC cord test).
Tested appliance must be turned on.
Test circuit is electrically isolated from the mains and from mains' PE lead.
Test result should be read only after displayed values are stabilized.
After the measurement the tested object is automatically discharged.
R
ISO
measurement in Class I appliances
Connect the mains plug of the tested appliance into the test socket of the tester
. The
measurement is made between L and N (which are shorted) and PE. Additionally, it is pos-
sible to carry out the measurement with the probe connected to
T2
terminal socket
.