PAT-80
● PAT-85 ● PAT-86 – USER MANUAL
50
PRCD cord (built-in RCD)
Connect the tested object to the meter.
Press
START
.
Upon completing the measurement read the result. You can finish the test
before the defined test time duration by pressing
STOP
.
Positive test result: I
T
≤ LIMIT
Negative test result: I
T
> LIMIT
NOTE!
During the measurement, the same mains voltage is present at the measuring
socket which powers the tested appliance.
During the measurement of a faulty appliance, RCD switch may be triggered off.
When
Change polarity
is set on
Yes
, after the set time duration is over the tester
automatically changes the polarity of the test mains socket and resumes the test.
As a test result it displays the value of the higher leakage current.
When tested appliance is powered from other socket, the measurement should be
performed at both mains plug positions and as the result the higher current value
should be accepted. When the appliance is powered from the
tester’s socket in auto
tests, L and N terminals are swapped by the tester.
The bandwidth of test current results from the measuring system with adjusted
touch current which simulates human perception and reaction, in accordance with
EN 60990:2002.