PAT-80
● PAT-85 ● PAT-86 – USER MANUAL
57
4.13 Power test
The test consists on measuring the power parameters when the tested appliance is turned on. The
powers, voltage, current are measured as well as cosφ, PF and THD factors.
Press
Power test
.
Before starting the measurement, similarly to
sec. 4.2
, following parameters must be set:
test dura-
tion
and
clamp
.
In
Use clamp
select:
Yes
if clamp is to be used
for the measurement,
No
if the test is to be per-
formed in test socket of
the tester.
Measurement with clamp
Attach clamp around L conductor. To
T1
socket connect L and N conductors of the power
cord of the tested appliance.