8-32
Rapidlab 1200 Operator’s Guide: System Configuration
02087462 Rev. V
Setting up Auto Send Options
You can select the type of results to automatically send to a connected LIS, Rapidlink, or
Rapidcomm data management system at the end of analysis.
•
Patient results
•
Calibration and QC results
Both types of results have a default value of On. Results are also sent when you select the
Print or the Send button.
1. At the Setup screen, select
Printer and Devices > Communications
.
NOTE:
You must select either RS-232 or CompleNet to make the Auto Send options
available.
2. Select
Patients
or
Calibrations and QC
to turn the option on or off.
3. Select
Save
.
Setting Up Remote Viewing (Rapidcomm Only)
NOTE:
Remote viewing and control require the cooperation of the Rapidcomm
administrator and operators at local systems. We recommend you establish a policy for the
remote viewing and control feature.
The display at local systems can be viewed and controlled remotely using the Rapidcomm
software. The Rapidcomm software enables the user at an LIS to monitor and troubleshoot
the local system remotely.
•
Before a remote user can view and control the local system, the local system operator
must enable the remote viewing feature and allow a remote user to connect to the local
system. This ensures that no conflict occurs between local and remote users who want
to view and control the system at the same time.
•
If a local system operator allows a remote viewer to view the local system, the remote
user can either view and take control of the local system or allow the operator to retain
control while the remote user only views the system display.
Enabling Remote Viewing
1. At the Setup screen, select
Printer and Devices
.
Select the down arrow and then select
Remote Viewer
.
At the Remote Viewer screen a Services column and a Configure column display.
2. To enable remote viewing, select
Start
in the Service column.
To disable remote viewing, select
Stop
in the Service column.
Stop is the default setting.
3. Select
Continue
.
Содержание Rapidlab 1200
Страница 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Страница 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Страница 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Страница 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Страница 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Страница 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Страница 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Страница 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Страница 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Страница 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...