Rapidlab 1200 Operator’s Guide: Maintenance
5-53
02087462 Rev. V
Flushing the Lower Connector
1. Connect the syringe tubing to the lower connector for the wash reagent.
2. Flush with a minimum of 3 mL water.
3. Flush with a minimum of 12 mL air.
4. Capture the water in an absorbent swab or equivalent.
Figure 5-27 Reagent Manifold Lower Connector
1
Water collection area
2
Lower connector
Содержание Rapidlab 1200