6-18
Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
2. Replace the
p
O
2
sensor with the clot clearing tool.
a. Push the black spring-loaded latch assembly (item 1, Figure 6-4), located at the
right of the 1200 Reference sensor, to the right, until the mechanism clicks in
place.
Figure 6-4 Measurement Module Compartment
b. Remove the
p
O
2
sensor by lifting the sensor lip (item 2, Figure 6-4) up at the
bottom while pressing the top of the sensor (item 3, Figure 6-4) inward.
1
spring-loaded latch assembly
2
top of the
p
O
2
sensor
3
lip of
p
O
2
sensor