6-24
Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
e. Press the black spring enclosure assembly lever to reapply pressure to the sensor
assembly.
10. Close the measurement module cover.
11. Close the front cover of the system.
12. Select
Stop
.
13. Select the Return button twice to exit Diagnostics mode.
You are prompted to calibrate the system.
14. Select
Yes
to calibrate the system.
15. Replace the reagent sample port by following this procedure:
a. Select
Status > Maintenance > Replace Sample Port
.
b. Select the
Continue
.
Follow the instructions in the video to replace the sample port.
16. If the reagent manifold, preheater, and reagent cartridge clot clearing procedures fail
to locate the obstruction, see
Removing Clots Using the Clot-Removal Line
Removing Clots Using the Clot-Removal Line
Use the following procedure if the clot clearing procedures in the preceding section fail to
remove the obstruction.
BIOHAZARD:
Wear personal protective equipment. Use universal precautions.
Refer to
Appendix A, Protecting Yourself from Biohazards
for recommended
precautions when working with biohazardous materials.
CAUTION:
Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the
measurement module frame to discharge static buildup.
Inspecting the Sample Path
1. Remove the reagent cartridge and the wash cartridge:
a. Select
Status > Diagnostics > Cartridges > Eject Both Cartridges
.
b. Remove both the cartridges.
c. Leave the doors open.
Содержание Rapidlab 1200
Страница 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Страница 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Страница 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Страница 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Страница 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Страница 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Страница 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Страница 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Страница 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Страница 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...