Rapidlab 1200 Operator’s Guide: Troubleshooting
6-69
02087462 Rev. V
D50 Glucose Sensor Error
Problem
Corrective Action
The system detects
an open connection
in the glucose
biosensor.
1. Check that the biosensor is correctly installed, the contacts
are aligned, and the spring-loaded latch is closed.
2. If the biosensor has been on the system less than one day,
perform three 2-point calibrations.
Refer to
Performing Manual Calibrations
3. If the problem persists, replace the biosensor.
4. If the problem persists, call for technical assistance.