Rapidlab 1200 Operator’s Guide: Maintenance
5-41
02087462 Rev. V
4. Add fill solution:
5. Install the internal reference electrode into the internal electrode compartment
and screw the electrode into place.
6. Clean and inspect the sensor:
a. Tap the front face of the sensor with your knuckle to remove any bubbles.
b. Wipe any excess fill solution from the exterior of the sensor using a
lint-free tissue.
c. Verify that the O-ring is in place on the left side of the sensor.
Replace the O-ring if it is worn or damaged.
7. Reinstall the sensor.
Refer to
8. Verify sensor performance.
Refer to
For these sensors...
Add...
pH
Na
+
, K
+
, Cl¯,
and Ca
++
K
+
, Cl¯, and Ca
++
sensors almost full but with a small
space (1 mm)at the top
CAUTION:
Do not touch the internal electrode wire. The wire is fragile and is
easily damaged.
Содержание Rapidlab 1200
Страница 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Страница 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Страница 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Страница 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Страница 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Страница 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Страница 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Страница 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Страница 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Страница 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...