
Rapidlab 1200 Operator’s Guide: Troubleshooting
6-83
02087462 Rev. V
User Canceled
Calibration
The operator interrupted a
calibration to perform another
action, for example to load a
STAT test.
No action necessary.
W Cartridge
Expired
The wash cartridge has exceeded
its operating life or is depleted.
Replace the cartridge. Refer to
W Cartridge
Not Valid
An expired wash cartridge was
installed or the cartridge was
installed after being removed for
more than 6 hours.
The system is unable to use the
cartridge.
Replace the cartridge. Refer to
Waste Full
The waste bottle is full and can
no longer accept waste.
Replace the waste bottle. Refer to
Waste is
Almost Full
The waste bottle is
approximately 70% full.
Replace the waste bottle. Refer to
page 2-2.
You can replace the waste bottle
now or wait until the system
prompts you.
Message
Probable Cause
Corrective Action
Содержание Rapidlab 1200
Страница 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Страница 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Страница 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Страница 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Страница 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Страница 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Страница 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Страница 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Страница 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Страница 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...