4-2
Rapidlab 1200 Operator’s Guide: Quality Control
02087462 Rev. V
c. Attach a Quick
®
adapter to the ampule.
d. Insert the adapter with the attached ampule into the sample port as shown on the
screen.
e. Select
Analyze
.
The system aspirates the sample.
f. When prompted, remove the Quick adapter from the sample port and then select
Continue
.
3. If prompted, enter your operator ID and then select
Continue
.
4. View the results on the screen and in the printed report.
If you turned on the Auto Print option in Setup, the system automatically prints the
QC sample report.
5. When you finish viewing the results, select
Continue
.
Procedural Notes for Required QC
When you perform a Required QC sample analysis, the system verifies that you scanned
the correct ampule for the scheduled control material. The system checks the results
against the ranges defined for each parameter for that level of the control. The system
identifies any results that are out of range.
The system allows you to analyze Unscheduled QC samples in addition to scheduled
Required QC analysis, but the results of the Unscheduled QC have no affect on parameter
status.
Depending on the options you selected in Setup, the system prints a QC sample report. If
the system is connected to a Rapidlink or Rapidcomm data management system or to an
LIS, the system automatically sends the QC results to these computer systems.
If...
Then...
you are prompted to
repeat analysis because
1 or more parameters is
out of range,
select
Yes
and repeat this procedure starting at step 2.
You can select
No
if you do not want to repeat the analysis.
The system returns to the Analysis screen and turns off the
parameters that failed QC analysis. You cannot report results
for these parameters until an authorized operator restores the
parameters. Refer to
1 or more parameters is
out of range for the
second time
the system returns to the Analysis screen and turns off the
parameters that failed QC analysis.
You cannot report results for these parameters until an
authorized operator restores the parameters. Refer to
Содержание Rapidlab 1200
Страница 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Страница 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Страница 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Страница 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Страница 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Страница 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Страница 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Страница 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Страница 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Страница 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...