2-14
Rapidlab 1200 Operator’s Guide: Operating the System
02087462 Rev. V
NOTE:
You can scan the patient ID and accession number in the appropriate field. For
information about entering patient sample data, refer to
8. View the results.
For information about how results are displayed on the screen and in printed reports,
refer to
9. Select
Continue
when you finish viewing the results.
Analyzing Microsamples
NOTE:
Read this procedure completely before performing microsample analysis for the
first time.
Use this procedure to analyze patient samples when you know that you have insufficient
sample volume for routine analysis. The procedure allows the system to have time to
properly analyze a small volume sample.
The system analyzes the sample in 2 stages. First, the system analyzes
p
CO
2
and
p
O
2
, and
then analyzes the remaining parameters. The system also measures
p
CO
2
during the
second stage of the microsample if the
p
CO
2
sensor still contains sample. Using the
following procedure, you position the sample at position 1 in the sample path to begin
analysis of
p
CO
2
and
p
O
2
. For Rapidlab 1245 and 1265 systems, the microsample mode
does not analyze hemoglobin parameters.
BIOHAZARD:
Wear personal protective equipment. Use universal precautions.
Refer to
Appendix A, Protecting Yourself from Biohazards
for recommended
precautions when working with biohazardous materials.
1. Observe all rules and guidelines in
2. At the Analysis screen, select the button for the patient sample type (arterial, capillary,
venous, mixed venous).
3. Select
Microsample
.
4. Scan the barcode for the patient sample if required.
5. Insert the sample device into the sample port and select
Analyze
.
CAUTION:
If the wash that is performed after sample analysis does not
completely clean the fluid path, you should perform an additional wash. To
determine if the path is clean, observe if any colored fluid remains in the fluid
path after analysis. To perform a wash at the Analysis screen, select
Wash
.
The Wash button is grayed out if the system is busy. The Wash button is not
available if the lamp failure or LIS communications error indicators, which
occupy the same space on the screen, are active.
Содержание Rapidlab 1200
Страница 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Страница 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Страница 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Страница 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Страница 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Страница 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Страница 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Страница 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Страница 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Страница 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...