8 - 1
Siemens-Elema AB
Register 3
SP B7-230.033.07
Page 1 of 2
Mammomat 1000/3000
Solna, Sweden
Rev. 03
11.99
SPS-UD
Installation and Start-Up
Checking the microprocessors
8
Microprocessors
8
The following LED displays on the processor P.C. boards indicate whether the relevant
microprocessors are operating correctly.
1.
Mains and system "ON".
The following conditions must be indicated after not more than 15 s.
P.C. board D711:
• V9 "ON" (line voltage on).
• V13 "OFF" (system stand-by).
• V11 "ON" (system on).
• V16 "ON" (intermediate circuit - filament).
MAM00110
Microprocessors for:
LED displays
Master: P.c. board D702
IONTOMAT PM: P.c. board D701
Must be off.
Flashing rapidly.
Must be off.
Must be flashing
(2-3 Hz).