![Siemens MAMMOMAT 3000 Скачать руководство пользователя страница 45](http://html.mh-extra.com/html/siemens/mammomat-3000/mammomat-3000_installation-and-start-up-instructions-manual_380661045.webp)
Mains connection and power supply
6 - 3
Siemens-Elema AB
Register 3
SP B7-230.033.07
Page 3 of 4
Mammomat 1000/3000
Solna, Sweden
Rev. 03
11.99
SPS-UD
Installation and Start-Up
Checking the supply voltages
6
The correct supply vo5 V, +15 V and +24 V have already been tested in the fac-
tory. Only a visual check of the function is therefore needed.
1.
System "ON".
2.
Check that the following orange diodes are "ON" on p.c. board D704:
V39 (+15 V)
V38 (-15 V)
V41 (+5 V)
V40 (+24 V)